FXT modes
Einstein Probe's Follow-up X-ray Telescope (FXT) performs prompt follow-up observations of sources detected by WXT and pointed observations of other selected targets. Table 1 summarizes the FXT parameters.
| FXT parameters | Value |
|---|---|
| Energy range | 0.3–10 keV |
| Mirror | Wolter-I |
| Detector | PN-CCD |
| Effective area | 600 cm2 at 1.25 keV (on-axis) |
| Field of view | 60' × 60' |
| Detector elements | 384 × 384 |
| Pixel scale | 75 μm × 75 μm |
| HPD | < 30″ (on axis) |
| Sensitivity | 2.4 × 10-13 erg cm-2 s-1 (1500 s exposure time) |
FXT can select a readout mode automatically according to the source count rate, or use a mode commanded from the ground. Full Frame (FF) and Partial Window (PW) modes retain two-dimensional imaging; Timing (TM) mode trades one spatial dimension for higher time resolution. Table 2 gives the timing properties and source-brightness limits. These limits depend on the filter and the acceptable pile-up fraction.
| Mode | Image capability (x×y) | Readout timing | Flux level (thin filter, pileup<2%) | Fraction of OoT |
|---|---|---|---|---|
| FF | 384×384, 2D | 50 ms | < 2 mCrab | 0.23% |
| PW | 128×61, 2D | 2.2 ms | < 40 mCrab | <1% |
| TM | 128×384, 1D | 23.68 μs (Read time per row) | < 100 mCrab (spectral analysis) < 1 Crab (timing analysis) | - |
For all three modes, onboard processing subtracts the bias and applies the lower-level discriminator. Events are not reconstructed onboard. During ground processing, fxtpical applies gain and CTI calibration and assigns calibrated PI channels, while fxtgrade reconstructs split-pixel event patterns and assigns event grades. The bad-pixel and hot/flickering-pixel screening steps add the corresponding event flags.
The Full Frame mode
Full Frame mode records 20 frames/s over the full imaging area. Each 50 ms cycle includes 49.8848 ms of integration and 0.1152 ms of frame transfer. During the next exposure, the three CAMEX chips read the stored frame in parallel.
Photons registered during charge transfer produce out-of-time (OoT) events. They are assigned incorrect positions along the readout direction. Because gain and Charge Transfer Inefficiency (CTI) corrections depend on position, their PI values may also be incorrect, broadening spectral features. The OoT fraction is approximately 0.23% in FF mode (Table 2).
FXT also records events in the frame-store area while the imaging area integrates. These events are mainly produced by charged particles and are identified by fsaevt in their filenames. They can be used to estimate the instrumental background in the imaging area.
The Partial Window mode
Partial Window (PW) mode reads a central 128×61-pixel region through the middle CAMEX (CAMEX_M). Each 2.2 ms cycle includes approximately 2.1319 ms of integration and 0.0681 ms of transfer. Figure 1 shows the readout region.

The readout window covers rows 162–222. Transferring the frame takes 227 row shifts in 0.0681 ms, or about 0.3 μs per shift. Each shift accounts for a fraction r=(0.3 μs)/(2200 μs)≈0.01364% of the frame cycle. For a source inside the window, the estimated OoT fraction is f≈(222-N)r, where N is the source Y coordinate (162–222).
Photons from outside the window can also be recorded during transfer. For sources in rows 0–161 or 380–383, the estimated OoT fraction is 61r≈0.832%. For rows 223–318 it is zero, and for rows 319–379 it is approximately (N-318)r.
The Timing mode
Timing mode is used for bright sources and high-time-resolution measurements. Rows are read consecutively at approximately 23.68 μs per row. After 384 rows, the readout node resets for about (6×23.68+200) μs, or 342.08 μs. Figure 2 shows the readout region.

Source events are spread along the readout direction, with an enhancement at the source position caused by photons collected during the reset interval.
The recorded frame time and readout order at CAMEX_M are used to reconstruct each event's arrival time during ground processing.
In the timing mode, there is an inherent periodic signal with a period of approximately 9.4352 ms (∼ 105.986 Hz). Figure 3 illustrates the profile of this periodic signal. It arises from the readout cycle and event-time reconstruction. Apparent variability near this frequency should be checked for an instrumental origin.

Charge loss during transfer depends in part on the incident flux, so the energy-channel relation can change at high count rates. For TM spectral analysis, the recommended source flux is below 100 mCrab; spectra above this level may have additional systematic errors.